Skip to main content Skip to secondary navigation

Scanning Preparation #3: Sample Dimensions

Main content start

Overview

Below is a basic outline of the ideal and maximum allowable dimensions of a sample for use in the UniTOM HR. More detailed descriptions and considerations for different sample dimensions are available by the provided links in the table or the side-menu. 

Sample Parameter

Recommended Configuration

       System Limit       

Sample Geometry

Cylindrical (no straight edges)

--

Sample Diameter

Less than 110 mm

500 mm

Sample Height

Less than 250 mm

700 mm

Sample Weight

--

45 kg

Further Considerations for Deben Stage

Below are the recommended sample dimensions for tensile and compression testing. 

Configuration

Recommended Sample Dimensions (in mm)

Tensile
Dimensions for samples used in the Deben stage during tensile loading.
Compression
Dimensions for samples used in the Deben stage during compression loading.